Provides the analysis of sample sizes typically in the 10 to 200-milligram range with the samples heated by a small furnace (~20 cc volume) capable of operating from 15 to 1600 degrees Celsius. The STA sensor is a double pan differential temperature sensor which is calibrated to generate data for heat flow to the sample with an accuracy of 5% or better. The weight sensor is located remotely below the furnace where it is isolated from sample decomposition products by inert purge through a narrow channel. This provides microgram - level weight change detectability of sample loss or oxidative gain.
Model: Thermal Analyzer STA 8000
Mass Spectrophotometer Clarus SQ 8 T
Brand: Perkin Elmer Inc.
Equipment Measurements & Capabilities:
1 | Simultaneous analysis of TG with DTA mode and DSC mode for fast enhanced result interpretation. |
2 | Vertical displacement balance sensor provides weight insensitivity to sample position, no spurious weight change from melting. |
3 |
|
Current Status of the Instrument : Active for TG DTA Analysis, MS under Maintenance
Location of the Instrument: Sahyadri Campus, Dr.APJ Abdul Kalam Block, CIF Lab, Ground Floor
Sample Submission Form (Internal users)
Sample Submission Form (External uses)