Provides the analysis of sample sizes typically in the 10 to 200-milligram range with the samples heated by a small furnace (~20 cc volume) capable of operating from 15 to 1600 degrees Celsius. The STA sensor is a double pan differential temperature sensor which is calibrated to generate data for heat flow to the sample with an accuracy of 5% or better. The weight sensor is located remotely below the furnace where it is isolated from sample decomposition products by inert purge through a narrow channel. This provides microgram - level weight change detectability of sample loss or oxidative gain.

https://lh3.googleusercontent.com/_zmsDtRHxtspXeJ3Y7WQVQ2LpEKEyOA0BKsqDIETsH7dt_1RumL0WjcwWvezgy73-LbXgrlEGrbH7t6dgvybUBRgH_nMQpzZkES0OV4KwdhtJEMHrRl9PWu_gdFOrU68VszebOLt

Model:  Thermal Analyzer     STA 8000

             Mass Spectrophotometer    Clarus SQ 8 T              


Brand: Perkin Elmer Inc.


Equipment Measurements & Capabilities:

1 Simultaneous analysis of TG with DTA mode and DSC mode for fast enhanced result interpretation.
2 Vertical displacement balance sensor provides weight insensitivity to sample position, no spurious weight change from melting.
3
  • Equipped with Mass analyser through Hyphenated technology enables Measurement of gases evolved from samples being tested by a thermal analyser. The Specifications of MS
  • Ionisation Type : Electron Ionisation
  • Signal to Noise ratio : 800: 1
  • Turbomolecular pump capacity : 255L/sec

 

 

Current Status of the Instrument : Under Maintenance

Location of the Instrument: Room No: 25, Ground Floor, Temporary Campus, IIT Palakkad

 

Sample Submission Form (Internal users)

 

Sample Submission Form (External uses)